8 (of 12) IQ procedure ROXY, edition 2
Verified by (customer): ………………………….. Deviations (Y/N): ……
Comments:
Table II
Check In
conf.
Non conf.
ref. *
Section "installation guide" and "safety prac-
tices" in users manual(s) is noticed
O
Environmental conditions are in accordance to
recommendations in manual
O
ROXY EC system trigger cable is correctly
connected & configured to the start input of the
mass spectrometer.
O
All tubing & connections are leakage-free in-
stalled in accordance with the corresponding
LC connection kit installation drawing.
O
Purity of all mobile phase chemicals are HPLC
grade or better
O
Installation procedure is done in accordance to
user's manuals.
O